Sysmex XP-300 14-43
Technical Information
Revised October 2012
5) 2B High UD with 20%-frequency distribution width error
RBC, HCT, MCV, MCH, MCHC, and PLT parameters are flagged. Out of the particle
distribution width data, RDW-CV is flagged with RU and its data are output.
RDW-SD is flagged with DW and its data are not output. In this instance, MCV and MCHC are
out of Patient Mark Limits.
• RBC Histogram Error - RU, DW (2B)
6) 2C High UD with 2 or more peaks
RBC, HCT, MCV, MCH, MCHC, and PLT parameters are flagged with RU. As to particle
distribution width data, RDW-CV is flagged with MP and its data are output, while RDW-SD is
also flagged with MP but its data are not output.
7) 3 20%-frequency particle distribution width error
RBC, HCT, MCV, MCH, MCHC, and PLT parameters are not flagged. As to the particle
distribution width data, RDW-CV data are output without flagging, while RDW-SD data are
flagged with DW and their data are not output.
8) 4 2 or more peaks
RBC, HCT, MCV, MCH, MCHC, and PLT parameters are not flagged. As to the particle
distribution width data, RDW-CV data are output with MP flag. RDW-SD is flagged with MP
and its data are not output. In this instance, RDW-CV is out of Patient Mark Limits.
• RBC Histogram Error - MP (4)
RBC RU 3.57 [×10
6
/µL]
HGB 11.5 [g/dL]
HCT RU 52.9 [%]
MCV RU +148.2 [fL]
MCH RU 32.2 [pg]
MCHC RU -21.7 [g/dL]
RDW-SD DW ---.- [fL]
PLT RU 185 [×10
3
/µL]
RBC 4.80 [×10
6
/µL]
HGB 14.2 [g/dL]
HCT 41.5 [%]
MCV 86.5 [fL]
MCH 29.6 [pg]
MCHC -34.2 [g/dL]
RDW-CV MP +33.3 [fL]
PLT 216 [×10
3