Sysmex XP-300 14-45
Technical Information
Revised October 2012
1) 1A High frequency for LD
PLT, PDW, MPV, and P-LCR parameters are flagged with PL.
• Platelet Histogram Error - PL (1A)
2) 1B High LD with 20%-frequency distribution width error
PLT parameter is flagged with PL. Particle distribution width (PDW) is flagged with DW and its
data are not output.
MPV and P-LCR are flagged with PL and their data are not output.
3) 1C High LD with 2 or more peaks
PLT parameter is flagged with PL. Particle distribution width (PDW) is flagged with MP and its
data are not output.
MPV and P-LCR are flagged with PL and their data are not output.
4) 2A High UD
PLT, PDW, MPV, and P-LCR parameters are flagged with PU.
5) 2B High UD with 20%-frequency particle distribution width error
PLT parameter is flagged with PU. Particle distribution width (PDW) is flagged with DW and
its data are not output.
MPV and P-LCR are flagged with PU and their data are not output.
• Platelet Histogram Error - PU (2B)
PLT PL 211 [×10
3
/µL]
PDW PL 12.5 [fL]
MPV PL 9.3 [fL]
P-LCR PL -22.6 [%]
PLT PU 171 [×10
3
/µL]
PDW DW ---.- [fL]
MPV PU ---.- [fL]
P-LCR PU ---.- [%]