14-46 Sysmex XP-300
Technical Information
Revised October 2012
6) 2C High UD with 2 or more peaks
PLT parameter is flagged with PU. Particle distribution width (PDW) is flagged with MP and its
data are not output.
MPV and P-LCR are flagged with PU and their data are not output.
7) 3 20%-frequency particle distribution width error
PLT is not flagged. PDW, MPV, and P-LCR are flagged with DW and their data are not output.
• Platelet Histogram Error - DW (3)
8) 4 2 or more peaks
PLT is not flagged. Other parameters are flagged with MP and their data are not output.
• Platelet Histogram Error - MP (4)
PLT
302010 [fL]
PLT 91 [×10
3
/µL]
PDW DW ---.- [fL]
MPV DW ---.- [fL]
P-LCR DW ---.- [%]
PLT 155 [×10
3
/µL]
PDW MP ---.- [fL]
MPV MP ---.- [fL]
P-LCR MP ---.- [%]