Horizontal and acquisition setups Select a sampling mode
Sampling Mode S
election MSO/DPO5054 and M SO/DPO5034
Channels on
123or4
Time base 20 n
s/div
RT ET RT
10 ns/div ET/IT
ETIT
ET/IT
Sampling Mod
e Selection MSO/DPO5204 and M SO/DPO5104
Channels on
123or4
Time base 2
0ns/div
RT RT RT
10 ns/div
RT RT ET
5 ns/div ET/IT ET/IT ET/IT
xxx
The type of equivalent-time sampling the instrument uses is called random equivalent-time sampling.
Although it takes the samples sequentially in time, it takes them randomly w ith respect to the trigger.
Random sampling occurs because the instrument sample clock runs asynchronously with respect to the
input signal and the signal trigger. The instrument takes samples independently of the trigger position and
displa
ys them based on the time difference between the sample and the trigger.
What do you want to do next?
Go to the Horizontal Setup control window. (see page 177)
Return to the Acquisition Setup control window. (see page 181)
192 DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals