EasyManuals Logo

Agilent Technologies E6198B User Manual

Agilent Technologies E6198B
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #128 background imageLoading...
Page #128 background image
5-60 E6198B Switch/Load Unit User Manual
5 Using Load Cards and Loads
Varistor voltage @ 1.0mA DC: 300 Vdc Max
Maximum Clamping Voltage (8/20mS): 450 volts
Capacitance: 17 pF
Typical Transzorb
®
specifications are:
Breakdown voltage: 300V max at 1mA
Reverse Standoff voltage: 250 volts
Maximum reverse leakage: 5mA
Maximum peak pulse current: 5A
Maximum Clamping voltage: 400 volts
Maximum temperature coefficient: 0.110%/°C
Load and UUT Connections
Figure 5-42 shows how the Agilent E6178B Load Card is wired.
Connect loads between J11-J18 and J31-J38. Loads are external
to the load card and connected to the load card via cables.
Connectors J11 and J31 connect to Channel 1, connectors J12
and J32 connect to Channel 2, etc.
When configured as part of a standard Agilent system, P2 of the
load card is typically connected via cables to a mass
interconnect panel. User connections to the UUT are then made
from the mass interconnect panel. Refer to the appropriate
mass interconnect wiring guide for connection details.
You can also make UUT and power supply connections directly
to P2 as shown in Figure 5-42.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6198B and is the answer not in the manual?

Agilent Technologies E6198B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6198B
CategoryTest Equipment
LanguageEnglish

Related product manuals