A-8 E6198B Switch/Load Unit User Manual
A Switch/Load Unit And Card Specifications
Agilent N9377A 16-Channel Dual-Load Load Card Specifications
Parameter Specification
Switch topology 16 channels with 2 loads per channel, off-board loads
Path resistance from power bus to load card
connect
0.4
Ω maximum
Exclusive of load, with standard 0.05
Ω sense resistor
and 10 A slo-blo fuse installed. (5
Ω for load currents
<1A).
Load path relay operate/release time 10/5 msec maximum, 30 cps max
Peak voltage to Earth 60 V continuous
500 V maximum transient, non-switching (decays to
<60V in 200mS)
Load carry/switching current: 7.5 A maximum non-switching, continuous
Load carry peak current 15 A max., non-switching (<100 msec, <2% duty cycle)
Load switching power 150 Watts maximum, resistive load
Load switching voltage 60 V maximum
Minimum permissible load 10 mA, 5 V
Basic current sense accuracy with standard 0.05W
sense resistor
0.1% maximum
Basic current sense accuracy with differential
amplifier engaged (Vcm=0V)
0.2% + 30mA
Basic current sense accuracy with differential
amplifier engaged (Vcm=16V)
0.2% + 100mA
Isense differential amplifier 3 dB bandwidth 500 kHz typical
Isense differential amplifier slew rate 40 A/uS typical
Maximum Current Consumption 0.5 A at 5 V
1 A at 12 V

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6198B and is the answer not in the manual?

Agilent Technologies E6198B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6198B
CategoryTest Equipment
LanguageEnglish

Related product manuals