Safety and Regulatory Information 2
E6198B Switch/Load Unit User Manual 2-3
Environmental Conditions
The Automotive Electronics Functional Test System is designed
for indoor use only. Table 2-1 shows general environmental
requirements.
Before Applying Power
Verify that the product is set to match the available line voltage
and all safety precautions are taken. Note the external markings
of the instruments described in “Safety Symbols and
Regulatory Markings”.
WARNING
DO NOT REMOVE RACK PANELS OR INSTRUMENT COVERS. Operating
personnel must not remove any rack panels or instrument covers.
Component replacement and internal adjustments must be made only
by qualified service personnel. Products that appear damaged or
defective should be made inoperative and secured against unintended
operation until they can be repaired by a qualified service personnel.
WARNING
The protection provided by the Agilent TS-5000 system may be
impaired if the system is used in a manner not specified by
Agilent.
Table 2-1 Environment Requirements
Environment Conditions Requirements
Maximum Altitude 2000 meters
Temperature (Operation) 5 C to 40 C
Maximum Relative Humidity The test system is designed to operate
in the range from 5% to 80% relative
humidity (non-condensing).
°
°
CAUTION
This product is designed for use in Installation Category II and
Pollution Degree 2, per IEC 61010-1.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6198B and is the answer not in the manual?

Agilent Technologies E6198B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6198B
CategoryTest Equipment
LanguageEnglish

Related product manuals