2.8 HbA1c Calibration
HA-8190V 2-41
b Set the sample tubes in ports 1 to 3 of the calibration rack.
Use any of the ports 1 to 3.
Set barcodes in any order.
5 Load the calibration rack onto the sampler.
a Load the calibration rack onto the sampler.
See step 4 in “2.5.1. Preparing Samples” on page 2-21.
To read the standard values from the calibration information barcodes:
Skip to step 7.
To enter the standard values from the operator panel:
Go to step 6.
Empty sample tubes with calibration
information barcode label (x 2)
12345678910
NOTE:
When the “sample ID auto-reading function” is turned off, set sample tubes with the barcode labels facing to
the main body of the instrument so that barcodes are read correctly.
IMPORTANT:
Check again to make sure that the dummy samples and
standard solutions are set in the proper ports. Calibration
cannot be performed if the samples and solutions are set in
the wrong ports.
Ports 1 to 3: Calibration information barcode (when needed)
Ports 4 to 6: Dummy sample (whole blood)
Port 7: Low solution for the Fast mode
Port 8: High solution for the Fast mode
Port 9: Low solution for the Variant mode
Port 10: High solution for the Variant mode
REFERENCE:
If sample rack transportation is set to “loop transportation”:
Load the empty start rack and then calibration rack onto the sampler.
To perform HbA1c control measurements following calibration:
Load the calibration rack and then control rack onto the sampler.
* For the control rack to use, see “2.7. HbA1c Control Measurement” on page 2-33.
To perform normal measurement following calibration:
Load the calibration rack and then normal racks onto the sampler.
To perform HbA1c calibration following normal measurements:
Load the normal racks and then calibration rack onto the sampler.
Calibration
information
barcode
Example
Dummy
samples
12345678910
Fast mode Low solution
High solution
Variant mode Low solution
High solution

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1c and is the answer not in the manual?

Arkray ADAMS A1c Specifications

General IconGeneral
BrandArkray
ModelADAMS A1c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals