Chapter 4 Maintenance
HA-8190V 4-60
10 Drain distilled water from the tubes.
a Remove the bottle caps with nozzle from the aluminium packs.
• Place the nozzles on the gauze.
• Wipe the nozzles all the way to the base.
b Discard all of the distilled water in the aluminium packs.
c Drain fluid from the chambers.
• See step 5 on page 4-53.
• The distilled water will be drained from the tubes.
11 Store the hemolysis washing solution bottle.
a Add distilled water to a beaker (500 mL or more capacity).
b Remove the bottle cap with nozzle from the hemolysis washing
solution bottle.
c Wipe solution from the nozzle with gauge and place the nozzle in
the beaker.
• Wipe the nozzles all the way to the base.
d Attach the cap (that was originally on the bottle before opening)
to the bottle and tighten it securely.
e Store the bottle at a temperature between 3°C and 30°C, avoiding
direct sunlight.
12 Wash the tube for hemolysis washing solution.
a On the standby screen, touch , <7. Maintenance menu>, <5. Drain menu> and <4. Hemolysis washing
sol.> in that order.
• The tube for hemolysis washing solution will be washed.
• When washing stops automatically, the [Drain menu] screen will appear again.
b Touch <4. Hemolysis washing sol.> again.
• Wait until washing finishes.
c Touch to return to the standby screen.
NOTE:
Hold the aluminium pack by the hard plastic neck while discarding the distilled water from the pack.
Dry the aluminium packs thoroughly, then keep them in the accessory case.