4.6 Before/After Extended Periods of Disuse
HA-8190V 4-61
13 Drain distilled water from the tubes.
a Discard all of the distilled water from the beaker in which the bottle cap with nozzle for hemolysis washing solution
was placed.
b Touch <4. Hemolysis washing sol.>.
The distilled water will be drained from the tubes for hemolysis washing solution.
The [Drain menu] screen will appear again.
c Touch to return to the standby screen.
14 Wash the nozzles.
a Wash the nozzles for eluents and hemolysis washing solution with distilled water, and blot with gauze.
Wipe the nozzles all the way to the base.
b Wrap the bottle caps with nozzle in gauze, place them in the plastic bag and put the bag on the bottle tray.
The tubes and fluid level detection sensor cord can remain connected to the instrument.
15 Discard liquid waste.
a Discard liquid waste remaining in the bottle for liquid waste.
16 Turn off the main power.
a Press the standby switch to turn off the power.
b Press the side of the main power switch on the rear panel to
turn off the main power.
c Unplug the power cord from the outlet.
17 Clean the chambers.
a Remove and clean the chambers.
See steps 7 and 8 on pages 4-57 to 4-58.
NOTE:
If you store the nozzles without washing, fluid may crystallize and clog the nozzles.
HOME
Standby switch
Main power switch

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1c and is the answer not in the manual?

Arkray ADAMS A1c Specifications

General IconGeneral
BrandArkray
ModelADAMS A1c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals