Chapter 5 Troubleshooting
HA-8190V 5-32
d Touch .
Fluid pumping will start.
Allow liquid to overflow from the connection.
Wrap the tissue paper tightly around the end of the tube to avert
any splattering.
e When liquid begins to leak from the connection, touch .
Fluid pumping will stop.
f Securely tighten the push screw that was connected in step 3-b.
Blot up any liquid overflowing from the connection.
4 Connect the spare tube to the joint.
a Loosen the push screw of the spare tube to the joint.
b Wrap the connection between the tube and joint with tissue paper.
The tissue paper blots up any liquid that leaks during fluid
pumping.
c Touch .
Fluid pumping will start.
Allow liquid to overflow from the connection.
d Securely tighten the push screw that was connected in step 4-a while liquid is overflowing.
Check that the liquid leak stops.
Go to “ Completing maintenance” on page 5-33.
Pump
NOTE:
Slightly loosen the push screw if liquid does not overflow
from the connection.
Suspend
a
NOTE:
Slightly loosen the push screw if liquid does not overflow
from the connection.
Restart