Chapter 1 Before Use
HA-8190V 1-26
5 Attach the bottle caps with nozzle to the packs and bottle.
a Hold the eluent A pack by the hard plastic neck and remove the
cap from the pack.
b Insert the nozzle of the bottle cap A into the eluent A pack.
Tighten the cap securely.
c Install the eluent B pack.
• See steps a and b.
d Install the eluent CV pack.
• See steps a and b.
e Install the hemolysis washing solution bottle.
• See steps a and b.
f Check that the caps of the eluent packs and hemolysis washing solution bottle are tightened securely.
NOTE:
Do not hold the eluent pack by the soft aluminium bag.
Eluent may spill and damage the instrument.
REFERENCE:
Keep the cap in the accessory case. This cap should be reused when transporting the instrument, or if the
instrument is not to be used for extended periods of time.
NOTE:
Attach the bottle caps with nozzle to the packs and bottle
somewhere other than above the instrument. Liquid may
spill and damage the instrument.
IMPORTANT:
If the cap is loose, liquid condensation may rise due to evaporation, resulting in inaccurate measurement
results.
NOTE:
Do not tilt the packs or bottle to a large degree once the bottle caps with nozzle have been attached to the
pack and bottle. The air inlet on the cap may clog.