EasyManua.ls Logo

BEAMEX MC4 - Page 117

BEAMEX MC4
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrator Adjustment
MC4 User Guide 105
Notes.
The Input/Output pairs are saved as floating point number
percentage values so the transfer function may be used for any
function and range. The percentage values are proportioned to the
current range.
The available free memory defines how many custom transfer
functions and Input/Output pairs may be saved. If there is not
enough memory left to save a custom transfer function, MC4
displays an error message suggesting that you first free some
memory and then retry adding a custom transfer function.
MC4 approximates the values between entered input/output pairs by
calculating a curve between a pair of points that connects smoothly
to the adjoining curve. The entered Input/Output pairs need to be
points of a strictly increasing function.

Table of Contents

Related product manuals