EasyManua.ls Logo

BEAMEX MC4 - Page 144

BEAMEX MC4
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General
132 MC4 User Guide
You can pause a calibration run whenever you want by pressing
F3/
Menu
. As long as the menu is open, the calibration is paused.
You may also view the calibration window and still keep the cali-
bration paused by selecting the
Pause
option. Then the calibra-
tion starts by pressing F3/
Continue
. In the menu, you also have
the possibility to, e.g.
Undo
a point or
Skip
a point. More of the
menu available in the calibration window can be found in chapter
Calibration Window Menu on page 145.
3. When the calibration run is
ready, MC4 opens a win-
dow telling whether the cal-
ibration was Passed or
Failed. The found Maxi-
mum error and the Reject
limit for the calibrated in-
strument are also shown.
Press F3/
Ok
to view the re-
sults of the current calibra-
tion .
O k
4 . 0 8 0 20 . 0 0
E r r :
0 . 5 0
5
0
+
-
C a l i b r a t i o n F i n i s h e d
M a x i m u m e r r o r : 0 . 5 1 .
R e j e c t l i m i t : 1 . 0 0 % ( O u t p u t )
P A S S E D
4. The result window always opens with the numeric result page open
(see leftmost picture below). To see more calibration result related
data, press F2/
More
. The second of all in all 3 result pages is the
Calibration Info
page where you are able to enter
Calibrated By
and environment data. See also notes later on in this chapter.
C a n c e l M o r e S a v e . . .
P A S S E D
0 . 0 0
2 5 . 0 0
5 0 . 0 0
7 5 . 0 0
1 0 0 . 0 0
7 5 . 0 0
5 0 . 0 0
2 5 . 0 0
0 . 0 0
4 . 0 8 1 8
8 . 0 7 4 8
1 2 . 0 6 0 7
1 6 . 0 5 0 8
2 0 . 0 4 7 1
1 6 . 0 5 1 7
1 2 . 0 6 1 3
8 . 0 7 6 1
4 . 0 8 0 2
0 . 5 1
0 . 4 7
0 . 3 8
0 . 3 2
0 . 2 9
0 . 3 2
0 . 3 8
0 . 4 8
0 . 5 0
I n p u t O u t p u t E r r o r
C a n c e l B o t t o m S e l e c t
P A S S E D
0 . 0 0
2 5 . 0 0
5 0 . 0 0
7 5 . 0 0
1 0 0 . 0 0
7 5 . 0 0
5 0 . 0 0
2 5 . 0 0
0 . 0 0
4 . 0 8 1 8
8 . 0 7 4 8
1 2 . 0 6 0 7
1 6 . 0 5 0 8
2 0 . 0 4 7 1
1 6 . 0 5 1 7
1 2 . 0 6 1 3
8 . 0 7 6 1
4 . 0 8 0 2
0 . 5 1
0 . 4 7
0 . 3 8
0 . 3 2
0 . 2 9
0 . 3 2
0 . 3 8
0 . 4 8
0 . 5 0
I n p u t O u t p u t E r r o r
S a v e A s F o u n d R e s u l t s
S a v e A s L e f t R e s u l t s
S a v e a s B o t h R e s u l t s
R e j e c t / C a n c e l C a l i b r a t i o n
5. Save/reject the results by pressing F3/
Save
. Select one of the
options in the rightmost picture above.
6. After the results are saved/rejected, here are some possibilities
of how you may continue:
Start another calibration run by pressing F3/
Start
.
Adjust the instrument by selecting F1/
Menu
,
Adjust Instru-
ment
.
End the calibration selecting F1/
Menu
,
Close
.

Table of Contents

Related product manuals