Examples of Instrument Calibration
144 MC4 User Guide
Note.
If the prescan stops and displays an error message, try widening the
scan range(s). It is also advisable to have the scan range(s) sym-
metrical in relation to the actuation/deactuation points. Also check
the
Step response wait
setting among calibration procedure data.
For more information of
Step response wait
field, see page 168.
Hint!
These instructions can be adapted for any kind of switch. Change
the input signal connections/settings according to the instrument's
input signal.
When calibrating an electrical instrument and the input signal is gen-
erated with MC4, use this example as the source for the input signal
part.