Maintaining MC4's Instrument Database
154 MC4 User Guide
MENU ITEM
(cont.)
DESCRIPTION
Create New
A submenu opens with the following op-
tions:
•
Create New Instrument
, which
opens the instrument data windows
for adding a new instrument. More
of all instrument data windows and
the fields they contain in chapter
Editing Instrument Data on page
156. See also notes at the end of
this chapter.
•
Create New Branch
, which opens
a window where you may enter the
name of the new plant structure lev-
el and write some notes describing
it. The name of the new branch has
to be unique.
Calibration Mode
Settings
This option opens a window where you can
edit general calibration related settings:
•
Environment Pressure Unit
. Al-
lows you to select a unit for the envi-
ronment pressure that suits your
needs.
•
Error Graph ± Scaling
. With this
setting you can define general
"zoom" of the calibration window's
graph. The reference is the reject
error limit. The default value is ±5
times larger graph than what is the
instrument's reject error limit.
•
Allow 'Save as Both' for Saving
Calibration Results
. Depending
on your company policy you may or
may not allow saving the results of
one successful calibration Both as
As Found and as As Left. Check or
uncheck this option according to you
company policy.