Appendix 1, Technical Data
MC4 User Guide 195
Type S
(2
Full Range: -50 ... 1768
Sub Range Resolution 1 Year Uncertainty
(1
(±)
-50 ... 0 0.01 °C 1.0 °C
0 50 0.01 °C 0.7 °C
50 1500 0.01 °C 0.6 °C
0 ... 1768 0.01 °C 0.7 °C
Type E
(2
Full Range: -270 ... 1000 °C
Sub Range Resolution 1 Year Uncertainty
(1
(±)
-270 ... -200 0.01 °C
(3
-200 ... 0 0.01 °C 0.07 °C + 0.08 % rdg
0 600 0.01 °C 0.07 °C + 0.015 % rdg
600 1000 0.01 °C 0.026 % rdg
Type J
(2
Full Range: -210 ... 1200 °C
Sub Range Resolution 1 Year Uncertainty
(1
(±)
-210 ... -200 0.01 °C
(3
-200 ... 0 0.01 °C 0.25 °C
0 ... 1200 0.01 °C
Type K
(2
Full Range: -270 ... 1372 °C
Sub Range Resolution 1 Year Uncertainty
(1
(±)
-270 ... -200 0.01 °C
(3
-200 ... 0 0.01 °C 0.1 °C + 0.1 % rdg
0 1000 0.01 °C 0.1 °C + 0.02 % rdg
1000 ... 1372 0.01 °C 0.03 % rdg
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period. (k=2).
Uncertainty does not include reference junction uncertainty.
2)
IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3)
±0.02 % of thermovoltage + 4 µV