Appendix 1, Technical Data
196 MC4 User Guide
Type N
(2
Full Range: -270 ... 1300 °C
Sub Range Resolution 1 Year Uncertainty
(1
(±)
-270 ... -200 0.01 °C
(3
-200 -100 0.01 °C 0.02 % rdg
-100 0 0.01 °C 0.15 °C + 0.05 % rdg
0 750 0.01 °C 0.15 °C + 0.01 % rdg
750 ... 1300 0.01 °C 0.03 % rdg
Type T
(2
Full Range: -270 ... 400 °C
Sub Range Resolution 1 Year Uncertainty
(1
(±)
-270 ... -250 0.01 °C
(3
-250 ... -200 0.01 °C 0.7 °C
-200 ... 0 0.01 °C 0.1 °C + 0.1 % rdg
0 400 0.01 °C 0.1 °C + 0.01 % rdg
Type U
(4
Full Range: -200 ... 600 °C
Sub Range Resolution 1 Year Uncertainty
(1
(±)
-200 ... 0 °C 0.01 °C 0.15 °C + 0.1 % rdg
0 ... 600 °C 0.01 °C 0.15 °C + 0.01 % rdg
Type L
(4
Range Resolution 1 Year Uncertainty
(1
(±)
-200 ... 0 0.01 °C 0.13 °C + 0.07 % rdg
0 900 0.01 °C 0.13 °C + 0.02 % rdg
Type C
(5
Full Range: 0 ... 2315 °C
Sub Range Resolution 1 Year Uncertainty
(1
(±)
0 900 0.01 °C 0.4 °C
900 2000 0.01 °C 0.045 % rdg
2000 2315 0.01 °C 1.2 °C
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period. (k=2).
Uncertainty does not include reference junction uncertainty.
2)
IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3)
±0.02 % of thermovoltage + 4 µV
4)
DIN 43710
5)
ASTM E 988 - 96