Troubleshooting 7
7.2 Status light in the OPUS spectroscopy software
The status light in the OPUS spectroscopy software is located on the lower right end of
the interface. It indicates the status based on the currently active channel or measure
-
ment experiment loaded.
The OPUS status light can indicate different types of operating states.
LED Definition
Gray:
no spectrometer connected
Green:
spectrometer is connected and works properly
Yellow:
warning (e.g. service life of a spectrometer com-
ponent comes to an end, instrument test has
expired etc.)
i Measuring is still possible.
Red:
error (e.g. detector not cooled, spectrometer
malfunction, defective spectrometer component,
instrument test failed etc.)
i Measuring is not possible.
Table 7.1: OPUS status light