177
Bruker Optik GmbH IFS 125M User Manual
Data Communication 9
9.2.3 Optional configuration: spectrometer connected to network PC
• PC must be equipped with 2 network cards.
• Three sets of addresses must be provided:
- for the spectrometer
- for the PC network card OPTIC CONNECTOR (communication PC/spectrome-
ter)
- for the PC network card LAN (communication PC/network)
Spectrometer OPTIC CONNEC-
TOR network card
LAN network card
IP address 10.10.0.1 10.10.0.2 assigned by network
administrator
Subnet mask 255.255.255.0 255.255.255.252 assigned by network
administrator
Gateway 0.0.0.0
do not define when
using Windows XP
0.0.0.0
do not define when
using Windows XP
assigned by network
administrator
Table 9.3: Network addresses when connecting spectrometer to network PC