A Specifications
Parameters Specifications
General Dimension Optics complete: 87 x 146 x 40 cm (w x d x h)
Optics without scanner: 87 x 67 x40 cm
High-resolution option: 87 x 216 x 40 cm
Weight Optics bench: 120 kg
Electronics unit: 35 kg
Power supply The spectrometer power supply unit automatically
adapts to the most common power sources.
• Valid voltage range: 100 V AC to 240 V AC
• Valid frequency range: 50 to 60 Hz
Data system:
120/230 V AC, 50 - 60 Hz (desktop)
100/240 V AC, 50 - 60 Hz (laptop)
(Installation complies with VDE 0100 or IEC)
Power consumption Basic spectrometer configuration: 120 W
Data system: 120 W (typical)
Overvoltage category
a
II, complies with 61010-1 or IEC 60664-1
Environmental condi-
tions
a
• Operational temperature range: 18°C - 35°C
Temperature variation: max. 1°C/ per hour and
max. 2°C per day
• Humidity (non-condensing): less than 80% (rela-
tive humidity)
• Installation site: in a closed room, max. 2000 m
above sea level
• Pollution degree: 2, complies with 61010-1 or IEC
60664-1
Safety regulations met • Complies with EN/IEC 61010-1/08.2002 (Safety
requirements for electrical equipment for
measurement, control and laboratory use)
• Complies with EN 60825-1/10.2003 (Safety
requirements for laser equipment)
EMI regulations met Complies with EN/IEC 61326/05.2004 (Electrical
equipment for measurement, control and labora
-
tory use - EMC requirements)
Table A.1: IFS 125M specifications