IFS 125M User Manual Bruker Optik GmbH
Installation 3
3.2 Packaging
3.3 Transportation
Due to its weight (approx. 120 kg) the spectrometer should be carried by at least 3 per-
sons. When transporting the spectrometer, use the original packaging to avoid dam-
ages.
3.4 Site requirements
The operating company has to provide an installation site that meets the following site
requirements:
CAUTION
Damaged packaging
Injury and/or property damage
Contact shipping company.
Do not put spectrometer into operation.
Environment: Temperature range: 18 - 35°C
Humidity (non-condensing): less than 80% (relative
humidity)
Electromagnetic inter-
ferences:
The spectrometer is shielded against electromag-
netic interferences.
Do not locate the spectrometer near potential induc-
tive electrical interference, high energy pulses and
high electric or magnetic fields.
Use lines which are separated from those which are
connected to machines (e.g. elevators, large electric
motors, etc.) with large electrical surges or bursts.
Table 3.2: Site requirements