IFS 125M User Manual Bruker Optik GmbH
Troubleshooting 7
7.3 Spectrometer problems indicated by an error message in
OPUS
Error messages which refer to the spectrometer can be displayed in OPUS as follows:
•in the Instrument Status dialog
by means of a yellow message bubble at the lower right end of the OPUS inter-
face
in a normal message window
i
In case of a spectrometer problem indicated by an error message, the OPUS status light
is yellow or red.
Error message in
OPUS
Possible cause Troubleshooting
End of average life-
time is nearly
reached, spare part
will be required.
The end of the specified life-
time of a spectrometer com-
ponent has nearly reached.
Contact Bruker service.
i Measuring is still possible.
Temperature out of
range.
Temperature inside the spec-
trometer is too high.
Cause: Ambient temperature
is too high.
Operate the spectrometer only in the
ambient temperature range specified
(
18 - 35°C) by the manufacturer.
Laser is off or no laser
signal.
Defective laser Contact Bruker service.
Scanner initialization
mode.
This error message appears
when you try to start a mea
-
surement while the spec-
trometer is still initializing.
i Further messages may be
displayed. As they do not
represent a spectrometer
problem you can ignore
them.
Wait until initialization has been fin-
ished.
If an error message
appears which is not
listed above,.......
...contact Bruker service.
Table 7.3: Error messages in OPUS