26
IFS 125M User Manual Bruker Optik GmbH
Installation 3
3.7.2 Connecting accessories via CAN bus
3.7.3 Connecting spectrometer to purge gas supply line
The spectrometer has one purge gas inlet to purge the interferometer and detector com-
partment. This is only required if the spectrometer is operated in a humid environment.
Tools required and not included in the delivery content: stiff purge hose (PVC), outer
diameter: 6
mm.
1 Connect the CAN bus cable to the CAN bus
connector located at the mobile electronics
unit.
Table 3.7: Connecting accessories via CAN bus
1 Pull off the plug from the purge gas inlet.
2 • Insert one end of the hose into the purge
gas inlet (A).
• Connect the other end of the hose to your
purge gas supply line. The supplied gas
must be dry and oil-free (see chapter
5.11
for details).
• Control the purge gas flow by the purge gas
controller (B).
Table 3.8: Connecting spectrometer to purge gas supply line