26
IFS 125M User Manual Bruker Optik GmbH
Installation 3
3.7.2 Connecting accessories via CAN bus
3.7.3 Connecting spectrometer to purge gas supply line
The spectrometer has one purge gas inlet to purge the interferometer and detector com-
partment. This is only required if the spectrometer is operated in a humid environment.
Tools required and not included in the delivery content: stiff purge hose (PVC), outer
diameter: 6
mm.
1 Connect the CAN bus cable to the CAN bus
connector located at the mobile electronics
unit.
Table 3.7: Connecting accessories via CAN bus
1 Pull off the plug from the purge gas inlet.
2 Insert one end of the hose into the purge
gas inlet (A).
Connect the other end of the hose to your
purge gas supply line. The supplied gas
must be dry and oil-free (see chapter
5.11
for details).
Control the purge gas flow by the purge gas
controller (B).
Table 3.8: Connecting spectrometer to purge gas supply line
BA