6.7 Transporting spectrometer
When transporting the spectrometer you have to fix the scanner on the scanning arm,
remove the beamsplitter and demount the complete scanning arm. Mind the order
described in the following chapters.
i
Moving the spectrometer within a building can be done without demounting the scanning
arm.
6.7.1 Removing covers
1 Removing scanning arm cover:
Open the clips on each side.
Put your hand below the clip and push the
clip upwards.
Pull the clip towards you until you feel resis-
tance.
i All the clips have to be in their outermost
position, otherwise you get problems when
lifting off the cover.
2 Two persons are required to lift off the cover
of the long scanning arm, one at the front
and one at the back end.
Put your hands below the hangers.
Carefully lift off the cover from the scanning
arm.
Cover not lift off properly.
Risk of scanner damage.
Lift off the cover straight and far
upwards.
Table 6.20: Transporting spectrometer - Removing cover from scanning arm
NOTE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker IFS 125M and is the answer not in the manual?

Bruker IFS 125M Specifications

General IconGeneral
CategoryLaboratory Equipment
SoftwareOPUS
TypeFTIR Spectrometer
DetectorsDTGS, MCT
Light SourceTungsten-Halogen
Interferometer TypeMichelson interferometer
VacuumOptional
Computer InterfaceEthernet

Related product manuals