Data Communication 9
9.3 Assigning network addresses
When delivered, the spectrometer uses the standard IP address 10.10.0.1. Assigning a
different IP address only becomes necessary if the spectrometer is to be directly con
-
nected to a network. This IP address has to be assigned by the network administrator.
The FCONF (Firmware Configuration) program is used to assign a new IP address to
the spectrometer. This program is part of the OPUS software. The single steps required
for assigning IP addresses are described in the following:
1 Start FCONF program.
☞ Double click the fconf.exe file in the OPUS base path.
2 • Activate the Assign a new address to
the spectrometer button.
• Click the Next button.
Table 9.4: Assigning network address