IFS 125M User Manual Bruker Optik GmbH
Firmware Update E
E.3 Performing firmware update
Figure E.1: FCONF program - Update firmware
1 Activate the Enter custom address
option button.
Enter the IP address in dotted nota-
tion.
i In case of a stand-alone spectrome-
ter configuration, the default IP
address is 10.10.0.1. If the spec-
trometer is directly integrated into a
network, it must have a different IP
address. In this case use the IP
address labeled on the front side of
the mobile electronics unit. The
operating company is obliged to
write the IP address on the said
label provided.
2 Check whether the requested spectrometer is recognized by the firmware.
Click the Beep button. If the spectrometer beeps three times in rapid suc-
cession (once in case of ALPHA), it has been recognized by the firmware.
Click the Next button.
Table B.1: Updating firmware