Specifications A
Signal exit port Option:
1 beam exit port for coupling an external detector
Detector options • Liquid N
2
-cooled MCT
• Liquid N
2
-cooled InSb
•DLATGS
• Silicon diode
• A variety of other high performance detectors
• Up to 2 internal detectors may be mounted per-
manently
Electronics Automation Remote control of the following devices:
• aperture wheel with 10 positions
• selection between external input and calibration
source
Options:
• up to 3 optical filter wheels with 8 positions each
• switching between 2 detectors
• 1 beam exit port and additional aperture wheel for
second detector
Amplifier Preamplifier with up to 4 gain levels, main amplifier
with 5 gain levels
A/D converter 24 bit Delta-Sigma ADC
Interface to host computer Ethernet connection 100Base-T with TCP/IP
Laser IFS 125M is a CLASS 2 LASER PRODUCT
Options Dual-channel acquisition Simultaneous data acquisition from 2 detectors or
from a 2-color detector
Accessory Solar tracker Computer-controlled solar tracker system for
atmospheric absorption spectroscopy
a. The values indicated are target values which have an effect on the performance of the entire configuration. In case of non-
observance the functioning and safety of the configuration may be affected adversely.
Parameters Specifications
Table A.1: IFS 125M specifications