IFS 125M User Manual Bruker Optik GmbH
Maintenance 6
4 To make the gain setting changes come into
effect you have to reset the controller. Alterna
-
tively, you can re-establish the contact to the
beamsplitter by the electronics as follows:
Carefully lift off the interferometer compart-
ment lid.
Open lid:
Risk of exposure to laser light, which can
cause permanent eye damage.
Do not stare into the beam.
Open the beamsplitter looking door once.
This is necessary to make the gain setting
changes come into effect.
5 If the signal for one beamsplitter is too low (even with high gain setting) while
the others have good signal, this particular beamsplitter has to be replaced.
Details on changing a beamsplitter are described in chapter 5.8.
6 If the signal is too low for all beamsplitters while the optical alignment is cor-
rect, the laser power itself may have dropped or one of the optical compo-
nents in the laser beam may have been contaminated. The fiber patch cable
could be broken or misaligned. Contact Bruker service.
Table 6.1: Quick check in OPUS
CAUTION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker IFS 125M and is the answer not in the manual?

Bruker IFS 125M Specifications

General IconGeneral
CategoryLaboratory Equipment
SoftwareOPUS
TypeFTIR Spectrometer
DetectorsDTGS, MCT
Light SourceTungsten-Halogen
Interferometer TypeMichelson interferometer
VacuumOptional
Computer InterfaceEthernet

Related product manuals