Bruker Optik GmbH IFS 125M User Manual
Maintenance 6
9 Pull the knob to the open position in order to
open the dewar evacuation valve and evacu
-
ate the dewar.
10 If the desired vacuum has been achieved,
gently push the knob to the closed position
in order to close the dewar evacuation valve.
Purge the section between vacuum pump
and vacuum adapter.
11 Rotate the knob several turns counterclock-
wise in order to screw the union nut off the
dewar evacuation valve. Be careful not to
purge the dewar.
12 Pull the knob to the open position.
13 Loosen the coupling nut and remove the
vacuum adapter from the connecting piece
of the dewar.
Reinstall the MCT detector into the spec-
trometer (chapter 5.10)
14 Check whether a signal is detected and the optics works properly. More
details are described in chapter
5.6.
Table 6.16: Evacuating detector