Instructions for Use Fabius MRI SW 3.n 13
For Your Safety and that of Your Patients
WARNING
When moving the anesthesia machine,
remove all monitors and equipment from the
top shelf and hinged arms, remove the
absorber system, vaporizers, and reserve gas
cylinders, push in the writing tray, and use
only the machine handles to push or pull the
unit (see “Accessory Mounting” on page 30).
The anesthesia machine should only be
moved by people who are physically capable
of handling its weight. Dräger recommends
that two people move the anesthesia machine
to aid in maneuverability.
Exercise special care so that the machine
does not tip when moving up or down
inclines, around corners, and across thresh-
olds (for example, in door frames and eleva-
tors). Do not attempt to pull the machine over
any hoses, cords, or other obstacles on the
floor.
WARNING
The Fabius MRI and its patient connections
must be carefully positioned so that the
patient cannot be disconnected when being
removed from the MRT system.
WARNING
To avoid the risk of electric shock, this equip-
ment must only be connected to a supply
mains with protective earth.
WARNING
Use only non-rebreathing systems having a
reservoir bag in compliance with ISO 5362
and/or a pressure relief valve to ensure cor-
rect pressure at the patient connection port. If
disregarded, this could lead to personal
injury.
CAUTION
Risk of pinching fingers or breathing hoses and
objects falling down
If the writing table is not engaged correctly,
objects can fall down and fingers or breathing
hoses can be pinched.
CAUTION
Do not use the following non-rebreathing sys-
tems: Magill, Kuhn or Bain. They are not MR safe
or MR conditional.
Otherwise the diagnostic quality of the MRT
images will be affected.
CAUTION
Use MR safe or MR conditional non-rebreathing
systems. Otherwise the diagnostic quality of the
MRT images will be affected.
CAUTION
Provide sufficient length of hoses for movement.
CAUTION
Communications with external equipment may be
temporarily affected by electromagnetic interfer-
ence due to the use of electrosurgical equipment.
CAUTION
Watch hoses connected to the patient while mov-
ing the Fabius MRI.
CAUTION
Only Vapor 2000 vaporizers can by used on the
Fabius MRI in MRT scanner rooms.
CAUTION
Risk of injury and image artifacts
The Fabius MRI is designed for use in MR envi-
ronment only as a system. The user should not
assume that individual components of the system
can be safely used in MR environment.
CAUTION
Risk of injury
Do not service this device while it is in the MR
environment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min