EasyManua.ls Logo

Gamatronic power+ sa - Calibrating AC Voltage; Figure 151: Enter Measured Value; Figure 152: Module Config. Setup Menu; Figure 153: Calibrate INPUT or OUTPUT Voltage

Gamatronic power+ sa
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GAMATRONIC ELECTRONIC INDUSTRIES LTD.
POWER
+
SA: 20, 30 and 40 kVA, Release 2.1
87
3. Adjust the measured value using the arrow keys, and press Ent to update:
[Main Menu > SETUP > (password) > Module conf. > DC Calibration > Ent]
MODULE V. CALIB – 15 Sec to expire
####### Press ENTER To UPDATE ######
MEASURED VALUE: 432.0V
Figure 151: Enter measured value
9.9.8 Calibrating AC Voltage
1. Select AC Calibration:
[Main Menu > SETUP > (password) > Module conf.]
1. Num of phase 5.Output Adjust
2. Module/s frequency 6.Frequency Limits
3. Module/s voltage 7.DC Calibration
4. Update Vo/Fr/ph 8.AC Calibration
Figure 152: Module Config. setup menu
2. Select either Calibrate INPUT Voltage or Calibrate OUTPUT Voltage and then
press Ent:
[Main Menu > SETUP > (password) > Module conf. > AC Calibration]
--- MODULES AC VOLTAGE CALIBRATION ---
SELECT – 1: Calibrate INPUT Voltage
SELECT – 2: Calibrate OUTPUT Voltage
Then press ENTER to continue
Figure 153: Calibrate INPUT or OUTPUT Voltage
3. Adjust the measured value using the arrow keys, and press Ent to update:
[Main Menu > SETUP > (password) > Module conf. > AC Calibration > Ent]
MODULE V. CALIB – 15 Sec to expire
####### Press ENTER To UPDATE ######
MEASURED VALUE: 240.0V
Figure 154: Enter measured value

Table of Contents

Related product manuals