Page #105 background imageLoading...
Page #105 background image
Color
Color on/off.
Flow Profiles
select between 3 different buttons
UtA
,
UmA
and
MCA
. Only one button can be
active at the same time. With pressing the button again it will be deactivated. Use
Flow Presets
for optimizing vessel flow representation scanning.
Info
Also see
'B-Mode'
on page 6-4 and
'Button description'
on page 3-5.
Cine Menu
For more information see
'Cine Mode'
on page 7-12.
6.2.4 M-Mode
M-Mode is intended to provide a display format and measurement capability that represents
tissue displacement (motion) occurring over time along a single vector.
M-Mode is used to determine patterns of motion for objects within the ultrasound beam. The
most common use is for viewing motion patterns of the heart.
Using M-Mode
1. Touch 2D on the user interface to start B-Mode.
2. Touch M on the user interface to start M-Mode.
3. The
M Main
menu appears .
4. Place the cursor line over the region of interest.
5. Press
2D/M run
(right or left trackball button).
6. Press
Freeze
.
Hint
To change M Gain rotate the 2D-button. To change the Frequency use the touch panel.
Info
Also see
'Button description'
on page 3-5. For the option AMM (Anatomical M-Mode) please
see
'Anatomical M-Mode (AMM)'
on page 6-12.
Note
The availability of some functions or features depends on the probe and ultrasound system
used.
M-Mode Main Menu
Display Format Different options for display arrangement.
AMM Opens the AMM mode.
Dyn. Contr. Dynamic Contrast
controls how echo intensities are
converted to shades of gray, thereby increasing the
adjustable range of contrast.
Harm. Frequ.
Multi Frequency mode lets you downshift to the probe's
next lower frequency or shift up to a higher frequency.
Cine #
Select the desired Cine #.
M-Mode Menu
Display Format Different options for display arrangement.
AMM Opens the AMM mode.
Dyn. Contr. Dynamic Contrast
controls how echo intensities are
converted to shades of gray, thereby increasing the
adjustable range of contrast.
2D Mode
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4 6-11