2.11 Guidance and manufacturer´s declaration
Guidance and manufacturer´s declaration - electromagnetic emissions
The Voluson™ SWIFT / Voluson SWIFT+ is intended for use in electromagnetic environment specified below. The
customer or the user of the Voluson™ SWIFT / Voluson SWIFT+ should assure that it is used in such an environment.
Emission test Compliance electromagnetic environment - guidance
CISPR 11/EN55011
Conducted and
radiated RF
EMISSIONS
Group 1 The Voluson™ SWIFT / Voluson SWIFT+ uses RF energy
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in
nearby electronic equipment.
Class A The Voluson™ SWIFT / Voluson SWIFT+ is suitable for
use in all establishments other than domestic, and may be
used in domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes,
provided the following warning is heeded:
Warning
: This equipment/system is intended for use by
healthcare professionals in a professional healthcare
environment only. This equipment/system may cause radio
interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the Voluson™
SWIFT / Voluson SWIFT+ or shielding the location.
IEC/EN 61000-3-2
Harmonic emissions
Class A
IEC/EN 61000-3-3
Voltage fluctuations/
flicker emissions
Complies
Note: The EMISSION characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR
11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment
might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
Guidance and manufacturer´s declaration - electromagnetic immunity
The Voluson™ SWIFT / Voluson SWIFT+ is intended for use in electromagnetic environment specified below. The
customer or the user of the Voluson™ SWIFT / Voluson SWIFT+ should assure that it is used in such an environment.
Immunity Type Equipment Capability Regulatory Acceptable Level Electromagnetic
environment - guidance
IEC/EN 61000-4-2
Electrostatic discharge
(ESD)
± 6 ,± 8 kV contact Edition 3
± 6 kV contact
±2, ±4, ± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
±2, ±4, ±8, ±15 kV Edition 4
±8 kV contact
±2, ±4, ±8, ±15 kV air
IEC/EN 61000-4-4
Electrical fast transient/burst
± 2kV for power supply lines
5 kHz and 100 kHz
repetition frequency
Edition 3
± 2kV for power supply lines
5kHz repetition frequency
±1 kV for input/output lines
5kHz repetition frequency
Mains power quality should
be that of a typical
commercial and/or hospital
environment.
±1kV for input/output lines
5 kHz and 100 kHz
repetition frequency
Edition 4
±2 kV for power supply lines
100 kHz repetition frequency
±1 kHz input/output lines
100kHz repetition frequency
IEC/EN 61000-4-5
Surge Immunity
± 0,5 kV, ± 1kV line to line Edition 3 and 4
± 0,5 kV, ± 1kV line to line
Mains power quality should
be that of a typical
commercial and/or hospital
environment.
± 0,5 kV, ± 1 kV, ± 2 kV Line
to earth
± 0,5 kV, ± 1 kV, ± 2 kV Line
to earth
Safety
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4 2-21