Display modes:
Simultaneous dual mode: 2D/2D+PD
Triplex mode: 2D+PD/PW
PD coding steps: 256 color steps
PD window size:
lateral: maximum to minimum B mode scan angle
axial: B-scan range
Display mode: P (power)
Wall motion Filter: 8 steps (low1, low2, mid1, mid2, high1, high2, max1, max2)
Smoothing Filter:
rising edge: 12 steps
falling edge: 12 steps
Gain control: +15dB to -15dB, 0.2dB steps
PD Ensemble: 7 to 31
PD Line Density: 10 steps
Pulse repetition frequency: 100 Hz to 20.5 kHz
PD Map: 8 different color codes for each probe
Flow Resolution: 4 steps (low, mid1, mid2, high)
Balance: from 25 to 225 in 41 steps
Artefact suppression: yes
13.10.7 HD-Flow™ Mode (HDF):
Screen formats: 2D+HDF (single, dual, quad)
Display Modes
Simultaneous dual mode: 2D/2D+HDF
Triplex mode: 2D+HDF/PW; 2D/M+MHDF
HD-Flow™ Coding Steps: 256 color steps
HD-Flow™ window size:
lateral: maximal to minimal B mode scan angle
axial: B-scan range
Display mode: H (HD Flow)
Wall Motion Filter: 8 steps (low1, low2, mid1, mid2, high1, high2, max1, max2)
Smoothing Filter:
12 steps rising edge;
12 steps falling edge;
Gain Control: +15dB to -15dB, 0.2dB steps
HD-Flow™ Ensemble: 7 to 31
HD-Flow™ Line Density: 10 steps
Pulse Repetition Frequency. 100Hz to 20.5KHz
HD-Flow™ Map: 8 different color codes for each probe
Flow Resolution: 4 steps (low, mid1, mid2, high)
Balance: from 25 to 225
Artefact suppression: yes