Adjust the Mix controls to use 50% of each mode. Rotate the Mix control to
change the percentage.
8.8.1.1 Render type: Gray
Gray Render Mode
In Gray Render Mode only the gray information of the data set is used, even if a Color Volume
image is displayed. In case of a data set without color information, this mode is automatically
activated.
Render Modes
Surface Smooth
The surface is displayed in a smoothed “texture” mode. The gray values of the surface are
identical with the gray values of the original scan.
HDlive™ Texture
Surface Texture
A surface will be displayed in “texture” mode. The gray values of the surface are identical with
the gray values of the original scan.
Max
The maximum gray values of the ROI are displayed.
Application: Representation of bony structures.
Min
The minimum gray values of the ROI are displayed.
Application: Representation of vessels and hollow structures.
X-Ray
Representation of the average gray values within the ROI.
Application: Tissue block with tumor or similar.
Table 8-6 Render Mode 1 (basic)
Surface Smooth
The surface is displayed in a smoothed “texture” mode. The gray values of the surface are
identical with the gray values of the original scan.
Light
A surface will be displayed in “light” mode. Structures close to the viewer are displayed bright;
structures more distant from the viewer are shaded.
The surface to be displayed has to be surrounded by hypo echoic structures (e.g. liquids).
Gradient Light
The surface will be displayed as if being illuminated from a spot light source.
The surface to be displayed has to be surrounded by hypo echoic structures (e.g. liquids).
Max
The maximum gray values of the ROI are displayed.
Application: Representation of bony structures.
Min
The minimum gray values of the ROI are displayed.
Application: Representation of vessels and hollow structures.
X-Ray
Representation of the average gray values within the ROI.
Application: Tissue block with tumor or similar.
Table 8-7 Render Mode 2 (mix)
Main Controls
Set Rendering
Select one of the following render groups:
Routine
,
HDlive™
,
Surface
,
Tissue
Bone
,
Heart
,
Select Presets
3D Contr.
Adjust the contrast.
3D Brightn.
Adjust the brightness.
Mix
Render mix mode: combine the two selected render modes and Gray Threshold low
for 3D images.
3D and 4D Mode
8-28
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson Swift and is the answer not in the manual?

GE Voluson Swift Specifications

General IconGeneral
BrandGE
ModelVoluson Swift
CategoryMedical Equipment
LanguageEnglish

Related product manuals