Window Length Range: 8 - 25; step size 1
Screen Formats
Single (2D/Elasto)
Dual (2D/Elasto+2D/Elasto)
Quad (2D/Elasto+2D/Elasto+2D/Elasto+2D/Elasto)
13.10.11 BFlow
Screen formats: single (BF), dual, (BF+BF), quad (BF+BF+BF+BF)
Display modes:
BF
Update: BF/PW
Acoustic Power range: 1 - 100
Scan angle: taken from 2D
GAIN range: +15 to -20 dB
Gray scale values: 8 bit
SRI taken from 2D
Persistence filter: 8 steps (pre)
S./PRI 1.00, 1.50, 2.00, 3.00, 4.00..................15.00
Quality: 3 steps (pre) low, norm, high
Enhance: 6 steps (pre) 0 - 5
Gray maps: 18
Tint maps: 10
Dynamic: 12 different dynamic curves C1 -C12
Accumulation: Off, 0.20, 0.35, 0.50, 0.75, 1.00, 1.50, Infinite
Background: 0, 1, 2
Frequency range 1-16 MHz ( depending on the probe, 3steps high, mid, low)
13.11 Generic Measurements and Measurements/Calculations
13.11.1 Generic Measurements
2D Mode and 3D: Distance: Distance (Point to Point), Distance (Line to
Line), 2D Trace (Trace Length&Point), Stenosis
(% Dist), Ratio D1/D2
Area/
Circumferen
ce:
Ellipse, Trace (Line & Point), Area (2 Dist)
Stenosis (% Area), Ratio A1/A2
Volume: 1 Distance, 1 Ellipse, 1 Dist. + Ellipse, 3
Distances
Angle: Angle (3 Point), Angle (2 Line)
M Mode: Generic Distance, Slope, Time, HR (Heart Rate),
Stenosis (% Dist)
Gen. Vessel IMT, Vessel Diam., Stenosis Diam., Time, HR
Technical Data/ Information