Language System
: Select the desired language and press
Save & Exit
. The system reboots by
itself, which is necessary to change the current language. Only languages available
on the system are listed. If a new language is installed, it is automatically added to
the list.
e-IFU
: Select the desired language for the e-IFU. This selection is not influenced by
the system language selection and vice versa.
Date Format
Select the desired date format.
Display
Select which of the following items should be displayed:
Show TGC curve
Screen Saver
when checked it is also possible to select
Activate System lock
with Screen Saver
Diverse
On/off selection for:
Auto Scan Stop
: If the system is not active, it activates freeze mode after 5 min
Note
The system activated freeze mode after 60 mins. even if the Auto Scan Stop is off.
Hardkey Beeper off
: The acoustic signal for pressing a hard key is turned off.
Softkey Beeper off
: The acoustic signal for pressing a soft key is turned off.
Show MI in orange > 1.0
: The MI at the Info Header must be colored in orange
if the MI is greater than 1.0. At reload no colored MI is displayed.
Limit MI to 1.0 in 3D/4D
: The MI is not higher than 1.0 if selected.
Display 2D Overlay Graphics
Display 3D/4D Rotation Sliders
TI Limitation
On/Off Selection for:
Activate TI Limitation
Show TI Limit notification
If GA is lower than 14w 0d, TI Limitation is available.
TI limit
: Choose between 0.7 and 1.0
VOLUSON
TM
Club Opens the
VOLUSON
TM
Club dialog.
Presentation Mode
Runs the presentation mode of the US-device.
Voluson™ Club
After clicking
Voluson™ Club
the following dialog appears:
Figure 11-3 Voluson™ Club dialog
The QR Image contains the URL to the Voluson™ Club registration and the serial number of
the Voluson™ SWIFT / Voluson SWIFT+ system.
Note
It may be necessary to install a QR App on your smartphone.
If you want to register later, please enter your email - address and click
SEND
.
System Setup
11-6
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson Swift and is the answer not in the manual?

GE Voluson Swift Specifications

General IconGeneral
BrandGE
ModelVoluson Swift
CategoryMedical Equipment
LanguageEnglish

Related product manuals