Figure 12-1 Battery Life cycle Graph 2
12.6.4.2 Starting Battery Mode
When the AC cable is unplugged or if there is an AC power failure, the system works with
Battery Power Mode. The system maintains the current operations and can scan or perform
other operations normally.
Note
The following message “Battery Mode” appears on the display when mode change from Power
Mode to Battery Mode or when program initialized.
12.6.4.3 Starting Power Mode
Note
The Following message "Power Mode" appears on the display when mode change from
Battery Mode to Power Mode or when program initialized.
12.6.5 Recovering System
In case of Battery Mode, the system does not display a “Battery Mode” message any more
when the system power is recovered.
12.6.6 Refresh the battery
To maintain the battery life, it is recommended to refresh the battery every 6 months. Refresh
procedure:
Turn on the system
Wait until the battery is fully charged. It takes at least 1 hour to fully charge the battery.
Turn off the system
Remove all probes
Turn on the system
Unplug the AC cable and wait until the system shuts down
Wait at least 5 hours
Plug in the AC cable
Turn on the system
Wait until the battery is fully charged
Peripheral Devices
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4 12-13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson Swift and is the answer not in the manual?

GE Voluson Swift Specifications

General IconGeneral
BrandGE
ModelVoluson Swift
CategoryMedical Equipment
LanguageEnglish

Related product manuals