Note
In case software version and user program version are not identical, one of the following
messages will be displayed:
When loading a Small Backup: "The current user programs are not compatible with this
software version".
When loading parts of a Small Backup: "The current user programs are not compatible
with this software version. Do you want to load the complete set of user programs?"
This can be solved by loading the latest available backup (Small Backup) which is compatible
with the software version.
Full System Configuration
The
Full System Configuration
backup contains the following data:
Patient demographic and exam data (database containing the patient data and
measurements)
Archive image data (NOT available when saving to the internal hard disk )
User Settings (databases and files containing gray curves and the user settings.)
Image transfer settings (DICOM settings e.g., DICOM servers, AE Title, Station Name,
etc.)
Measure Setup Settings (user specific measure settings)
Voluson™ SWIFT / Voluson SWIFT+ settings (general settings such as language, time/
date format and the enabled options)
Windows Network Settings (network settings including the computer name)
Service platform (state of the service platform)
VP (additional system data)
All settings and patient data created since last full system configuration backup are NOT
backed-up! It is highly recommended to create a full system configuration backup of settings
and patient data regularly.
Saving the Full System Configuration:
1.
Click
Save
and the Save window is displayed.
2. Choose the destination (e.g., Network Drive).
3. Enter a description.
4. If desired and possible, activate
Include Images
.
Note
This can be a large amount of data, up to 70 gigabytes!
5. Click
Next
and confirm with
Yes
to start the backup process.
6. After copying the data, confirm the next message with
OK
to reboot the system.
Remark
It is possible to store more than one backup on a destination. The backups reside in
subfolders of the main Fullbackup - folder found at the root of the drive (e.g.,
Z:\Fullbackup). DO NOT modify this directory structure or any files within, otherwise the
backup data cannot be restored. .
The
Include Images
” checkbox is only active, if destination
Network Drive
or
Other drive
is selected.
If the destination
Other drive
is selected, the available drives (e.g., external USB-memory
stick) can be chosen from the drop down list.
System Setup
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4 11-41

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson Swift and is the answer not in the manual?

GE Voluson Swift Specifications

General IconGeneral
BrandGE
ModelVoluson Swift
CategoryMedical Equipment
LanguageEnglish

Related product manuals