8.8.4 Speckle Reduction Imaging (SRI)
Speckle Reduction Imaging (SRI) can be activated to reduce speckle.
Note
If selected in System Setup - User Settings, SRI affects slices and rendered image. Therefore,
it is also active in Full Screen mode.
In addition, if SRI is activated in 2D mode, it is automatically activated in 3D/4D VolPre mode
and automatically affects the image after/during acquisition.
Caution
Filters smoothen the final image (structures may be smeared out). For diagnostic purposes,
the Region of Interest must be checked without filters. A smoothed image might lead to false
diagnosis!
Activate SRI and change the level of smoothing in the sectional planes using the
- +
keys. Use
of SRI is indicated in the info block.
8.8.5 Volume Cine
The 4D VolCine function allows the user to save and work with the acquired Volumes.
Depending on the memory and volume size up to volumes can be displayed. The advantage
of working with 4D VolCine is that during the acquisition the user can concentrate on the
acquisition itself. After acquisition the user has the possibility to review and work on the
acquired volumes.
3D and 4D Mode
8-34
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4