Caution
Cleaning agents can lead to skin irritations. Ensure that no residue of the cleaning agent
remains on any part of the system after cleaning.
Caution
It is highly recommended to create a full backup of settings and patient data regularly.
The data from the backup always replaces the corresponding data on the Voluson™ SWIFT /
Voluson SWIFT+.
Caution
Do not touch the patient and the signal input/signal output (SIP/SOP) connectors
simultaneously.
Caution
Be careful when adjusting mechanical parts of the equipment.
Ensure nothing is jammed.
Do not put your hands or other body parts between movable parts of the equipment.
Be careful when adjusting or locking the monitor.
Caution
Position the Voluson™ SWIFT / Voluson SWIFT+ so that the AC Mains circuit breaker and the
AC Mains plug at the Mains wall-outlet is easily accessible at all times.
Caution
Disconnect mains from AC wall outlet to make device electroless.
Caution
Do not stare into the light beam of the LED.
Note
GE ULTRASOUND KOREA, LTD. is not responsible for any damage caused by viruses,
malware and other harmful software.
Note
The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals, e.g. professional healthcare environment (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally required) this equipment
might not offer adequate protection to radio-frequency communication services. The user
might need to take mitigation measures, such as relocating or re-orienting the equipment.
No Type Cable Construction / Manufacturer Max. Length (m) Type of Cable
1 AC cable H05VV-F 3G1.0mm² Manufactured by VOLEX/
FELLER
4 Non-Shielded
2 AC cable RVV 3G1.0mm²
Manufactured by VOLEX/FELLER
4 Non-Shielded
3 AC cable VCTF 3G2.00mm2 Manufactured by VOLEX/
FELLER
4 Non-Shielded
4 AC cable SJT 3/14AWG
Manufactured by VOLEX/ FELLER
4 Non-Shielded
Table 2-1 Main Power Cable
Safety
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 R
evision 4 2-9

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson Swift and is the answer not in the manual?

GE Voluson Swift Specifications

General IconGeneral
BrandGE
ModelVoluson Swift
CategoryMedical Equipment
LanguageEnglish

Related product manuals