1
PASS hold function
This function retains PASS state to help verify the value screened in the test.
Selection
0: Not held (initial setting)
1: Held
2
FAIL hold function
This function retains FAIL state to help verify the value screened in the test.
Selection
0: Not held
1: Held (initial setting)
3
Hold function
Enable this function to hold the current state when testing is interrupted by the
STOP
key.
Selection
0: Not held (initial setting)
1: Held
4
Momentary-out
The momentary out function allows current output only while the
START
key
is held down.
The
START
key on the remote control or the START signal via external I/O
has the same effect.
Selection
0: Not set (initial setting)
1:Set
5
Double action
Enable this function to allow testing to start only when the
START
key is
pressed within about 0.5 seconds after the
STOP
key.
Selection
0: Not set (initial setting)
1:Set
6
FAIL mode
Enable this function to restrict hold release to the
STOP
key on the main unit.
Selection
0: Not set (initial setting)
1:Set
7
Interface command "START"
Turn this function on to enable the interface START command.
Selection
0: Not set (initial setting)
1:Set
8
Inter-lock function
Enable this function to activate the external I/O interlock terminals.
Selection
0: Not set (initial setting)
1:Set
9
Output-voltage restricting value
Enable this function to set the upper-limit value for the output voltage
(common for the withstand-voltage and insulation-resistance tests).
Selection
0.5 to 5.0 kV (Can be set in 0.1-kV steps) (initial setting: 5.0 kV)