173
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:MODE <data>
<data>
MWITH
: Withstand-voltage mode
MINS
: Insulation-resistance mode
AWI
: Withstand-voltage mode insulation-resistance mode
AIW
: Insulation-resistance mode withstand-voltage mode
PROG
: Program mode
(Character data)
Function
In the READY state, sets the test mode.
Example
Sets the test mode to the auto test mode (withstand-voltage mode insulation-resistance mode.)
Transmission :MODE AWI
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:MODE?
Response
syntax
<data>
<data>
MWITH
: Withstand-voltage mode
MINS
: Insulation-resistance mode
AWI
: Withstand-voltage mode insulation-resistance mode
AIW
: Insulation-resistance mode withstand-voltage mode
PROG
: Program mode
(Character data)
Function
Queries the test mode.
Example
Queries the test mode.
Transmission :MODE?
Response AWI
Test mode: withstand-voltage mode
insulation-resistance mode
:MODE
Sets the test mode.
:MODE?
Queries the test mode.