174
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:STATe?
Response
syntax
<data>
<data>
WPASS
: Withstand-voltage mode PASS
IPASS
: Insulation-resistance mode PASS
WUFAIL
: Withstand-voltage mode UPPER FAIL
IUFAIL
: Insulation-resistance mode UPPER FAIL
WLFAIL
: Withstand-voltage mode LOWER FAIL
ILFAIL
: Insulation-resistance mode LOWER FAIL
WREADY
: Withstand-voltage mode READY
IREADY
: Insulation-resistance mode READY
WTEST
: Withstand-voltage mode TEST
ITEST
: Insulation-resistance mode TEST
NULL
: Others(Aborts a test)
(Character data)
Function
Queries the state.
Example
Queries the state.
Transmission :STATe?
Response WREADY
In the withstand-voltage mode READY state.
Syntax
:STARt
Function
Starts a test in the READY state.
Note
To start a test using this command, set "START" for the interface command to "1:Set",whichis
an optional function setting (for details, see Section 7.7 Interface command "START").
Example
Starting a test
Transmission :STARt
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
In the optional functions, if the RS command START is set to "0: Not set," an execution error
will occur.
Note that an execution error occurs when double action is set to "1: ON" on the optional
function setting screen.
:STATe?
Queries the state.
:STARt
Starting a test.

Table of Contents