10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:STOP
Function
In the TEST state, forcibly ends a test.
Furthermore, releases the Hold function and returns to the READY state. However, in the
optional functions, when "FAIL Mode Function" is set to ON, the Hold function cannot be
disabled by this command. (See Section 7.6.)
Example
Forcibly ends a test.
Transmission :STOP
Error
In the optional functions, when "FAIL Mode Function" is set to ON, an execution error will
occur if the unit is in the FAIL Hold Mode.
Syntax
:CONFigure:WITHstand:KIND <data>
<data>
AC50,AC60,DC
(Character data)
Function
Sets the type of voltage for withstand-voltage tests in the READY state.
Example
Sets the type of voltage for withstand-voltage tests at AC 50 Hz.
Transmission :CONFigure:WITHstand:KIND AC50
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:CONFigure:WITHstand:KIND?
Response
syntax
<data>
<data>
AC50,AC60,DC
(Character data)
Function
Queries the type of voltage for withstand-voltage tests in the READY state.
Example
Queries the type of voltage for withstand-voltage tests.
Transmission :CONFigure:WITHstand:KIND?
Response AC50
The type of voltage is AC50 Hz.
:STOP
Forcibly ends a test and releases the hold state.
:CONFi
ure:WITHstand:KIND
Sets the type of test voltage.
:CONFi
ure:WITHstand:KIND?
Queries the type of voltage for withstand-voltage tests.