176
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:CONFigure:WITHstand:VOLTage <data>
<data>
0.20 to 5.00
(NRf)
Function
Sets the test-voltage value (unit: kV) for withstand-voltage tests in the READY state.
Example
Sets the test-voltage value for withstand-voltage tests at 1.00 kV.
Transmission :CONFigure:WITHstand:VOLTage 1.00
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:CONFigure:WITHstand:VOLTage?
Response
syntax
<data>
<data>
0.20 to 5.00
(NRf)
Function
Queries the test-voltage value (unit: kV) for withstand-voltage tests.
Example
Queries the test-voltage value for withstand-voltage tests.
Transmission :CONFigure:WITHstand:VOLTage?
Response 1.00
The test-voltage value is 1.00 KV.
Syntax
:CONFigure:WITHstand:CUPPer <data>
<data>
0.1 to 100
(AC-withstand-voltage test)
0.1 to 10
(DC-withstand-voltage test)
(NRf)
Function
Sets the upper-limit value (unit: mA) for withstand-voltage tests, in the READY state.
Example
Sets the upper-limit value for withstand-voltage tests to 5.0 mA.
Transmission :CONFigure:WITHstand:CUPPer 5.0
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
:CONFi
ure:WITHstand:VOLTa
e
Sets the test-voltage value for withstand-voltage tests.
:CONFi
ure:WITHstand:VOLTa
e?
Queries the test-voltage value for withstand-voltage tests.
:CONFi
ure:WITHstand:CUPPer
Sets the upper-limit value for withstand-voltage tests.