10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:CONFigure:WITHstand:CUPPer?
Response
syntax
<data>
<data>
0.1 to 100
(AC-withstand-voltage test)
0.1 to 10
(DC-withstand-voltage test)
(NR1 or NR2)
Function
Queries the upper-limit value (unit: mA) for withstand-voltage tests.
Example
Queries the upper-limit value for withstand-voltage tests.
Transmission :CONFigure:WITHstand:CUPPer?
Response 5.0
The upper-limit value is 5.0 mA.
Syntax
:CONFigure:WITHstand:CLOWer <data>
<data>
0.1 to 99
(AC-withstand-voltage test)
0.1 to 9.9
(DC-withstand-voltage test)
(NRF)
Function
Sets the lower-limit value (unit: mA) for withstand-voltage tests, in the READY state.
Example
Sets the lower-limit value for withstand-voltage tests to 0.1 mA.
Transmission :CONFigure:WITHstand:CLOWer 0.1
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:CONFigure:WITHstand:CLOWer?
Response
syntax
<data>
<data>
0.1 to 99
(AC-withstand-voltage test)
0.1 to 9.9
(DC-withstand-voltage test)
(NR1 or NR2)
Function
Queries the lower-limit value (unit: mA) for withstand-voltage tests.
Example
Queries the lower-limit value for withstand-voltage tests.
Transmission :CONFigure:WITHstand:CLOWer?
Response 0.1
The lower-limit value is 0.1 mA.
:CONFi
g
ure:WITHstand:CUPPer?
Queries the upper-limit value for withstand-voltage tests.
:CONFi
g
ure:WITHstand:CLOWer
Sets the lower-limit value for withstand-voltage tests.
:CONFi
g
ure:WITHstand:CLOWer?
Queries the lower-limit value for withstand-voltage tests.