178
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:CONFigure:WITHstand:TIMer <data>
<data>
0.3 to 999
(NRf)
Function
Sets the test time (unit: s) for withstand-voltage tests, in the READY state.
Example
Sets the test time for withstand-voltage tests to 30.0 s.
Transmission :CONFigure:WITHstand:TIMer 30.0
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:CONFigure:WITHstand:TIMer?
Response
syntax
<data>
<data>
0.3 to 999
(NR1 or NR2)
Function
Queries the test time (unit: s) for withstand-voltage tests.
Example
Queries the test time for withstand-voltage tests.
Transmission :CONFigure:WITHstand:TIMer?
Response 30.0
The test time is set to 30.0 s.
Syntax
:CONFigure:WITHstand:UTIMer <data>
<data>
0.1 to 99.9
(NRf)
Function
Sets the ramp-up time (unit: s) for withstand-voltage tests in the READY state.
Example
Sets the ramp-up time for withstand-voltage tests at 10.0 s.
Transmission :CONFigure:WITHstand:UTIMer 10.0
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
:CONFi
g
ure:WITHstand:TIMer
Sets the test time for withstand-voltage tests.
:CONFi
g
ure:WITHstand:TIMer?
Queries the test time for withstand-voltage tests.
:CONFi
g
ure:WITHstand:UTIMer
Sets the ramp-up time for withstand-voltage tests.