179
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:CONFigure:WITHstand:UTIMer?
Response
syntax
<data>
<data>
0.1 to 99.9
(NR2)
Function
Queries the ramp-up time (unit: s) for withstand-voltage tests.
Example
Queries the ramp-up time for withstand-voltage tests.
Transmission :CONFigure:WITHstand:UTIMer?
Response 10.0
The set ramp-up time is 10.0 s.
Syntax
:CONFigure:WITHstand:DTIMer <data>
<data>
0.1 to 99.9
(NRf)
Function
Sets the ramp-down time (unit: s) for withstand-voltage tests in the READY state.
Example
Sets the ramp-down time for withstand-voltage tests at 5.0 s.
Transmission :CONFigure:WITHstand:DTIMer 5.0
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:CONFigure:WITHstand:DTIMer?
Response
syntax
<data>
<data>
0.1 to 99.9
(NR2)
Function
Queries the ramp-down time (unit: s) for withstand-voltage tests.
Example
Queries the ramp-down time for withstand-voltage tests.
Transmission :CONFigure:WITHstand:DTIMer?
Response 5.0
The set ramp-down time is 5.0 s.
:CONFi
g
ure:WITHstand:UTIMer?
Queries the ramp-up time for withstand-voltage tests.
:CONFi
g
ure:WITHstand:DTIMer
Sets the ramp-down time for withstand-voltage tests.
:CONFi
g
ure:WITHstand:DTIMer?
Queries the ramp-down time for withstand-voltage tests.